VL-4000 Current Loop & Voltage Datalogger

Version shown begin shipping Fall 2010.

Request pricing & availability

Data Sheets

Generation 1 VL-4000 Data Logger [pdf]
A4 Version [pdf]

NEW VL-4000 Data Logger [pdf]
A4 Version [pdf]

 

Overview

XPS Excitation switch

Validatible Precision Current Loop (4-20mA) & Voltage Datalogger


Specifications for the Veriteq VL-4000 line of precision current loop (4-20mA) and voltage dataloggers.

General

Size 3.4 x 2.3 x 1" (85x59x26mm);
76g (2.7 oz)
Operating Range -40 to 85°C (-40 to 185°F) and
0 to100% RH (non-condensing)
Interfaces RS-232 serial port; USB; WiFi; Ethernet; PoE
Mounting Magnetic strips; 3M Dual Lock™ fasteners
PC Software Compatible with the following software packages: Veriteq vLog, Veriteq viewLinc (for on-going monitoring & alarming), Veriteq OPC Server (to allow your Veriteq data loggers to work with other systems)
Internal Clock Accuracy: +/- 1 min./month at -25 to +70°C (-13°F to +158°F)
Electromagnetic Interference FCC Part 15, CE EN 55022:2006, EN 610000-4-2:2001; EN 61000-4-3:2006
Power Source Internal lithium battery with life of 10 years (Battery life specified with sample interval of 1 min. or longer)

Current loop & voltage inputs

Input Type Current Loop Analog Voltage
Available Range(s) 0 to 22mA 0 to 5 VDC, 0 to 10 VDC
Resolution 5.5 microamps 0.025% F.S.
Accuracy +/-0.15% F.S. at +25°C (+77°F) +/-0.15% F.S. at +25°C (+77°F)
Input Impedances 75 ohms >1 MOhm
Isolation One common per logger One common per logger
Overload Protection 40mA max. (reverse-polarity protected) +/-24 VDC max. (reverse-polarity protected)

Memory

Memory Type Non-volatile EEROM
Data Sample Capacity 120,000 12-bit samples
Memory Modes User-selectable wrap (FIFO) or stop when memory is full. User-selectable start and stop times.
Memory Protection Data retention > 20 years backup without power.
Sampling Rates User-selectable (in 10 second intervals) from once every 10 seconds to once a day. (Battery life specified with sample interval of 1 min. or longer)
Recording Span Recording span depends upon sample interval selected and number of channels enabled. Chart below details typical sampling rates and length of time logger will retain data in memory before wrapping around or stopping (see Memory Modes).
Sample Interval
Recording Span
Number of channels
1 2 3 4
10 seconds 13.8
days
6.9
days
4.6
days
3.4
days
1 minute 2.7
months

1.3
months

27.7
days
20.8
days
5 minutes 1.1
years
6.9
months
4.6
months
3.4
months
15 minutes 3.4
years
1.7
years
1.1
years
10.4
months
1 hour 13.6
years
6.8
years
4.5
years
3.4
years