VL-1000 Validatable Temperature Logger

Request pricing & availability

Related Information

VL-1000/1400 Version 6 Datasheet (PDF) (8.5x11 format | A4 format)

VL-1000/1400 Version 5 Datasheet (PDF) (8.5x11 format | A4 format)

Validation System Overview

Validation System Components

VL-1000-VLT Very Low Temperature Logger

Data Logger Rentals

Request Veriteq's 21 CFR Part 11 White Paper

Validatable Temperature Logger - VL 1000/1400


  • Portable self-contained temperature data logger for critical pharmaceutical validation applications
  • Equipped with internal temperature sensor, memory and 10-year battery
  • Tamper-proof operation produces secure 21 CFR Part 11 compliant documents
  • Password-protected setup, download and calibration functions ensure integrity of collected data
  • Records temperature without wires, power cords or connections; immune to power failures and network interruptions
  • NIST-traceable instrument-grade temperature measurement accuracy
  • Adjustable recording intervals (every minute, every hour, etc) with multi-year data recording capacity
  • High-stability temperature sensor provides in-calibration performance between calibration intervals up to a year
  • External input for remote temperature probes

Technical Specifications

Size 3.4 x 2.3 x 1" (85x59x26mm);
76g (2.7 oz)
Operating Range

-40 to 85°C (-40 to 185°F)
0 - 100% HR non-condensing

Interfaces RS-232 serial,USB, Ethernet,WiFi, PoE network interface available; USB interface available through USB Logger Cable. Contact Veriteq for more information.
Mounting Magnetic strips; velcro optional
Software Designed for use with viewLinc and vLog software
Clock Accuracy ±1 min./month @ -25°C to +70°C (-13°F to +158°F)
Electromagnetic Interference Meets FCC Part 15 for digital devices; meets CE Requirements for radiated emissions, electrostatic discharge, & radiated susceptibility.
Power Source Internal lithium battery with life of 10 years at 1 min. sampling rate.

Temperature Recorder Accuracy

Sensor Calibrated
Measuring Range
Resolution One year Accuracy
Internal -25°C to +70°C 0.02°C at +25°C

±0.15°C over +20°C to +30°C

±0.25°C over -25°C to +70°C

“N” Range External
Probe
-25°C to +70°C 0.02°C at +25°C ±0.15°C over +20°C to +30°C

±0.25°C over -25°C to +70°C
“L” Range External Probe -50°C to +10°C 0.02°C at -20°C

±0.25°C over -50°C to +10°C

“V” Range External Probe -90°C to -40°C 0.02°C at -80°C ±0.25°C over -90°C to -40°C

Internal Temperature Sensor

Type of Sensor Precision-tolerance epoxy-encapsulated NTC thermistor
Response Time 10 seconds in moving air (thermistor only)

Memory

Memory Type Non-volatile 32K x 8 EEROM
Data Sample Capacity 1000-2XX - 1000-LT 48,100 12-bit samples
1400-44X 85,300 12-bit samples
Memory Modes User-selectable: Wrap when memory full, or Stop when memory full.
Memory Protection Data retention > 20 years backup without power.
Sampling Rates User-selectable (in intervals of 10 seconds) from once every 10 seconds to once a day. Battery life specified with a sampling rate of 1 per minute.
Recording Span Dependant upon the sample interval selected and number of channels enabled. The table below details typical sampling rates and length of time logger will retain data in memory before wrapping around or stopping (see Memory Modes, above).
1000 -2xx Sample Interval Recording Span
1 channel 2 channels
10 seconds 5.5 days 2.7 days
1 minute 1.1 months 16.7 days
5 minutes 5.5 months 2.7 months
15 minutes 1.3 years 8.3 months
1 hour 5.4 years 2.7 years
1400-44x   1 channel 2 channels 3 channels 4 channels
10 seconds 9.8 days 4.9 days 3.2 days 2.4 days
1 minute 1.9 months 29.6 days 19.7 days 14.8 days
5 minutes 9.8 months 4.9 months 3.2 months 2.4 months
15 minutes 2.4 years 1.2 years 9.8 months 7.4 months
1 hour 9.7 years 4.8 years 3.2 years 2.4 years