More information:Thermocouple Logger Overview Datasheet (371 KB PDF File, 4 Pages) See Data for other Probes:
|
Type J Thermocouple Performance DataThe following charts document the performance of Veriteq 1700 Thermocouple data loggers when used with Type J (Iron-Constantan) thermocouples. Type J Temperature Measuring Range
Type J Temperature Measurement AccuracyThe following graph details the instrument accuracy of the Veriteq 1700 Thermocouple data logger over the rated Type J measurement ranges for both Wide Range ("W") models (top graph line) and Narrow range ("N") models.
Instrument accuracy is based on a 25° C. cold junction temperature. Probe errors and probe temperature range limitations are excluded. About Type J ThermocouplesType J thermocouples are made up of a positive iron wire and a negative Constantan wire. They are one of the most popular thermocouple types primarily because of a broad measurement range and superior voltage output (which can result in greater temperature resolution). Note that type J probes, because of the iron component, are susceptible to oxidation and rusting, particularly at sub-zero temperatures. |
||||||||||||||||||