SP-1700 Thermocouple data logger

More information:

Thermocouple Logger Overview
[VL 1700] [SP 1700]

Common Questions

Datasheet (371 KB PDF File, 4 Pages)

See Data for other Probes:



Type J Thermocouple Performance Data


The following charts document the performance of Veriteq 1700 Thermocouple data loggers when used with Type J (Iron-Constantan) thermocouples.

Type J Temperature Measuring Range

1700 Datalogger Version Type J Measuring Range

(°C.)

(°F.)


Wide Range "W" models -50 to 900 -58 to 1650

Narrow Range "N" models -130 to 160 -200 to 320

Type J Temperature Measurement Accuracy

The following graph details the instrument accuracy of the Veriteq 1700 Thermocouple data logger over the rated Type J measurement ranges for both Wide Range ("W") models (top graph line) and Narrow range ("N") models.

Type J Temperature Accuracy Graph

Instrument accuracy is based on a 25° C. cold junction temperature. Probe errors and probe temperature range limitations are excluded. 

About Type J Thermocouples

Type J thermocouples are made up of a positive iron wire and a negative Constantan wire. They are one of the most popular thermocouple types primarily because of a broad measurement range and superior voltage output (which can result in greater temperature resolution). Note that type J probes, because of the iron component, are susceptible to oxidation and rusting, particularly at sub-zero temperatures.